Configuration
10.4.8
Utilities > Configuration > Test > FRC (MBW)
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
Test Procedure
You can configure whether the motor and valve unit closes automatically or
manually. If you have activated manual close, you need to choose the button
Close
to close the motor and valve unit during the DLCO test.
To configure that the motor and valve unit closes manually, select the check
box
Activate Manual
Close.
Simulator Test-Gas
For the DLCO calibration check, you can use a DLCO simulator. The DLCO
simulator requires a test gas mixture with a higher accuracy.
Enter the test gas concentrations as specified on the gas cylinders into the
fields
CO
and
O
. The test gas concentrations must be within the range
2
indicated in parentheses.
Volumes
Enter physiological dead space volumes that are used for the DLCO test into
the fields
Anatomic Dead Space
be within the range indicated in parentheses.
The manufacturer strongly recommends that you select the
Volume
check box to automatically determine the discard volume.
Deselecting the
Automatic Discard Volume
Parameter
To select which diagnostic parameters to include on the test result screens
and on the printed reports, choose Select.
Viewing and printing reports with EasyOne Pro/LAB, 102
Test Mode and Min. No. of Breaths
After the number of breaths specified in the
automatically or the
Activate
The number of breaths in the
parentheses.
For the automatic test start choose Automatic. For the manual test start
choose Manual.
Display Mode Tracer Graph
To display the tracer curve with a linear scaling, choose Linear.
To display the trace curve with a semi-logarithmic scaling choose Semi-Log.
and
System Dead
Space. The volumes must
check box can cause false results.
#
field, the test either starts
button is displayed to start the test manually.
#
field must be within the range indicated in
Automatic Discard
121
116/171