Electromagnetic Emission; Electromagnetic Immunity - NDD EasyOne Pro Operator's Manual

Hide thumbs Also See for EasyOne Pro:
Table of Contents

Advertisement

Appendix
A.4.5

Electromagnetic emission

Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
 Emission table for IEC 60601-1-2 3
A.4.6

Electromagnetic immunity

Immunity test standard
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
 Electromagnetic immunity
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
Guidance and manufacturer's declaration – electromagnetic
emissions
EasyOne Pro/LAB is intended for use in the electromagnetic environment
specified below. The user of EasyOne Pro/LAB should assure that it is used in
such an environment.
Compliance
Electromagnetic environment – guidance
Group 1
EasyOne Pro/LAB uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Class B
EasyOne Pro/LAB is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
Class A
buildings used for Harmonic emissions domestic purposes.
Complies
rd
th
and 4
edition
The following tables are guidelines according to the 3
medical standard IEC 60601-1-2.
Guidance and manufacturer's declaration – electromagnetic Immunity
EasyOne Pro/LAB is intended for use in the electromagnetic environment
specified below. The user of EasyOne Pro/LAB should assure that it is used in
such an environment.
IEC 60601 test level
Compliance level
±6 kV contact
±8 kV contact
±8 kV air
±15 kV air
±2 kV for power supply
±2 kV for power supply
lines
lines
±1 kV for input/output
lines
±1 kV line to line
±1 kV line to line
±2 kV line to earth
±2 kV line to earth
<5% U
(0.5 cycles)
<5% U
T
40% U
(5 cycles)
40% U
T
70% U
(25 cycles)
70% U
T
<5% U
for 5 s
<5% U
T
Note: U
is the AC mains voltage prior to application of the test level.
T
Electromagnetic environment – guidance
None
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
(0.5 cycles)
Mains power quality should be that of a
T
typical commercial or hospital
(5 cycles)
T
environment.
(25 cycles)
T
for 5 s
T
rd
edition of the
169/171
171