Appendix
Rated maximum output
power of transmitter
0.01 W
0.1 W
1 W
10 W
100 W
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Note: An additional factor of 10/3 is used in calculating the recommended separation distance to decrease the
likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought
into patient areas.
 Recommended separation distances between portable and mobile RF communications equipment and EasyOne Pro/
LAB
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
Recommended separation distances between portable and mobile RF
communications equipment and EasyOne Pro/LAB
EasyOne Pro/LAB is intended for use in the electromagnetic environment in
which radiated RF disturbances are controlled. The user of EasyOne Pro/LAB
can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(transmitters) and EasyOne Pro/LAB as recommended below, according to the
maximum output power of the communication equipment.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d
0.35 P
=
0.035 m
0.11 m
0.35 m
1.1 m
3.5 m
80 MHz to 800 MHz
d
0.35 P
=
0.035 m
0.11 m
0.35 m
1.1 m
3.5 m
800 MHz to 2.5 GHz
d
0.7 P
=
0.07 m
0.22 m
0.70 m
2.2 m
7.0 m
171/171