Servicing and troubleshooting

12
Servicing and troubleshooting
12.1

Checking for correct operation of EasyOne Pro/LAB

12.2

List of troubleshooting solutions

Problem
When the EasyOne Pro/LAB is switched
on, you see the error message that is
printed in the next column.
When you start a test, you see the
message Check spirette insertion.
Calibration check outside ±3.5%.
 Troubleshooting solutions
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
1 Perform a calibration check.
2 Perform a spirometry test on yourself or on another known subject.
3 Check the results of steps 1 and 2 for plausibility.
a) If the results of steps 1 and 2 are plausible, EasyOne Pro/LAB is
operating correctly.
b) If the results of steps 1 and 2 are not plausible, troubleshoot.
c) If you are not able to remedy the defect, contact your EasyOne dealer or
the ndd Servicing Department.
 Calibration check, 122
Performing lung function tests, 48
List of troubleshooting solutions, 127
Contact information, 4
If you encounter problems operating EasyOne Pro/LAB, consult the table
below.
Possible cause
Self-test failed
The spirette is not correctly positioned. Ensure that the triangle on the flow
The spirette is not correctly positioned. Ensure that the triangle on the flow
You have not used an ndd adapter.
There are leaks in the syringe
connection.
The specified syringe volume differs
from the actual syringe volume.
 List of quality messages and quality grades, 86
Contact information, 4
Solution
Quit the program and restart. If you
receive the same message again,
contact your EasyOne dealer or the
ndd Servicing Department.
sensor is lined up with the triangle on
the spirette.
sensor is lined up with the triangle on
the spirette.
Use the spirette cal check adapter.
Check the connections.
Choose the correct syringe volume
under
Configuration > Test > Cal-
Check.