Performing An Mbw Test - NDD EasyOne Pro Operator's Manual

Hide thumbs Also See for EasyOne Pro:
Table of Contents

Advertisement

Performing lung function tests
6.6.6

Performing an MBW test

Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
6 To return to the main menu, choose
 Cleaning instructions for EasyOne Pro/LAB, 99
About interpreting results, 85
Description of printer configurations of EasyOne Pro/LAB, 118
Required materials
A new spirette for each patient
A new FRC barriette for each patient
A nose clip
Prerequisites
You have added the patient data to the database of EasyOne Pro/LAB.
Or the patient data is already in the database of EasyOne Pro/LAB.
The patient has performed an SVC test.
The results of the SVC test are required for the interpretation of the
FRC test.
You have inserted the one-way valve (multi use).
You have inserted the FRC barriette (single-patient use)
You have mounted the motor and valve unit onto the flow sensor.
You have connected the gas supply and opened the gas valves.
You are wearing disposable gloves.
You have instructed the patient on how to perform the test before you
start with the test.
Adding patient data to the database of EasyOne Pro/LAB, 100
Performing the breathing maneuver for the SVC test, 97
The one-way valve for multi-patient use, 18
The DLCO barriette and FRC barriette for single-patient use, 19
Mounting the motor and valve unit onto the flow sensor, 65
Connecting the gas supply and opening the gas valves, 66
WARNING
PATIENT HEALTH HAZARD
Performing lung function tests can cause an asthma attack or bronchospasm.
If the patient is on medication, check the contraindications for the medica-
tion.
Main Menu
from the menu bar.
93
77/171