Performing lung function tests
6.6.6
Performing an MBW test
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
6 To return to the main menu, choose
Cleaning instructions for EasyOne Pro/LAB, 99
About interpreting results, 85
Description of printer configurations of EasyOne Pro/LAB, 118
Required materials
A new spirette for each patient
A new FRC barriette for each patient
A nose clip
Prerequisites
You have added the patient data to the database of EasyOne Pro/LAB.
Or the patient data is already in the database of EasyOne Pro/LAB.
The patient has performed an SVC test.
The results of the SVC test are required for the interpretation of the
FRC test.
You have inserted the one-way valve (multi use).
You have inserted the FRC barriette (single-patient use)
You have mounted the motor and valve unit onto the flow sensor.
You have connected the gas supply and opened the gas valves.
You are wearing disposable gloves.
You have instructed the patient on how to perform the test before you
start with the test.
Adding patient data to the database of EasyOne Pro/LAB, 100
Performing the breathing maneuver for the SVC test, 97
The one-way valve for multi-patient use, 18
The DLCO barriette and FRC barriette for single-patient use, 19
Mounting the motor and valve unit onto the flow sensor, 65
Connecting the gas supply and opening the gas valves, 66
WARNING
PATIENT HEALTH HAZARD
Performing lung function tests can cause an asthma attack or bronchospasm.
If the patient is on medication, check the contraindications for the medica-
tion.
Main Menu
from the menu bar.
93
77/171