Performing lung function tests
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
3 Tell the patient to exhale as hard and as fast as possible and to continue
blowing out until the lungs are completely empty.
4 When the message Manually End Test is displayed, you can choose the
Button
End Test
at your discretion.
 You can configure manual test stop or automatic test stop under
Utilities > Configuration > Test > General > Test
 The test result is displayed.
5 Tell the patient to take the spirette out of the mouth and to breathe
normally again.
 If the trial is acceptable, the green quality message is displayed.
 If the trial is not acceptable, a yellow or red quality message is
displayed suggesting how to improve the breathing maneuver.
 To add trials
1 Choose
Add
Trial.
2 Repeat the previous procedure To perform the FVC breathing maneuver
and this procedure To add trials until the green Session Complete!
message is displayed.
 In the
Test Information
3 Do the following:
a) Review the test quality that is displayed in the
b) Decide whether the test quality is sufficient or not.
c) If necessary, add a trial and repeat the previous procedure To perform
the FVC breathing maneuver.
 To end the test
1 Grip the spirette with the wrapper again and pull the spirette out of the
spirette holder. Do not touch the spirette.
2 Dispose of the spirette together with the wrapper.
Procedure.
area the quality grade for the test is displayed.
Test Information
area.
93
56/171