Performing lung function tests
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
CAUTION
PATIENT CROSS-CONTAMINATION
If you reuse the spirette, DLCO barriette, or FRC barriette, contamination from
an infected patient can be deposited on the spirette, on the DLCO barriette, or
on the FRC barriette. The contamination from an infected patient can later be
passed on to the next patient who is tested.
You cannot clean or disinfect the spirette, the DLCO barriette, or the FRC
barriette in any way. The spirette, the DLCO barriette, and the FRC barriette are
for single-patient use only.
Always replace the spirette, the DLCO barriette, and the FRC barriette with a
new one between patients or when performing lung function test on yourself.
Only use an original spirette, DLCO barriette, and FRC barriette by the manu-
facturer.
The spirette is a single-patient-use disposable and hygienically packaged.
Visually check the wrapper of a new spirette for defects.
If the wrapper of the spirette has a defect, discard the spirette, take another
spirette, and visually check the wrapper of the new spirette again for defects.