Specifications And Bibliography; List Of Specifications For Easyone Pro - NDD EasyOne Pro Operator's Manual

Hide thumbs Also See for EasyOne Pro:
Table of Contents

Advertisement

Specifications and bibliography

13
Specifications and bibliography
13.1

List of specifications for EasyOne Pro

Size
Weight
For the flow sensor
Measuring accuracy
Volume
Flow
MVV
Measuring range
Resolution
Flow
Volume
Resistance
For the gas sensors
Main gas sensor
CO
Range
Resolution
Accuracy
Tracer gas sensor
He
Range
Resolution
Accuracy
Molar Mass Analyzer
Resolution
Accuracy
Measurement principle
For spirometry
For DLCO
Sample rate
Input method
Test storage capacity
Respiratory tube
Languages
Printing option
Export/EMR
Hardware interface
 Specifications for EasyOne Pro/LAB
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
These specifications of the EasyOne Pro model differ slightly from the
specifications of the EasyOne Pro LAB model.
List of specifications for EasyOne Pro LAB, 133
27 x 33.5 x 27 centimeters (10.6 x 13.2 x 10.6 inches)
7.3 kilograms (16.09 pounds)
±2% or 0.050 L
±2% or 0.020 L/s, (except PEF) PEF: ±5% or 0.5 L/min
±2% or 0.050 L/min
±16 L/s
4 mL/s
1 mL
Approximately 0.3 cm H
0% to 0.35%
0.0001%
±0.001%
0% to 50%
0.02%
0.05%
0.005 g/mol
0.01 g/mol
Ultrasonic transit-time measurement
Non-dispersive infrared, ultrasonic transit-time measurement
400 Hz
Touchscreen
More than 100,000 tests
Disposable spirette respiratory tube
English, French, German, Spanish, Italian, Brazilian Portuguese, Dutch,
Russian, Vietnamese, Turkish, Chinese, Swedish
PCL standard, direct to printer or over network
HL7, XML, GDT, via USB, LAN Network
Ethernet port, USB, possibility to upgrade to Wi-Fi
O/L/s
2
137
130/171

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Easyone pro lab