Performing lung function tests
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
 To end the DLCO test
1 Grip the spirette with the wrapper around again and pull the spirette out of
the spirette holder. Do not touch the spirette.
2 Dispose of the spirette together with the wrapper.
3 CAUTION! Always wear disposable gloves. Between patients, always
exchange disposable gloves, clean the flow sensor, and disinfect hands.
Make sure that no fluid penetrates the flow sensor or the inside of
EasyOne Pro/LAB while cleaning. To clean the flow sensor, do not immerse
the flow sensor into any fluid.
To clean the flow sensor of EasyOne Pro/LAB and to disinfect your hands
after each patient, do the following:
a) Clean the outside of the flow sensor as described in the cleaning
instructions
Hygiene and cleaning,
b) Put down the flow sensor of EasyOne Pro/LAB.
c) Take off or change the disposable gloves and disinfect your hands
before you put on new disposable gloves.
4 To print the report immediately, choose
screen.
 The report is generated and sent to the printer that you have configured.
5 To preview the report, do the following:
a) Choose
Report
from the bottom right of the screen.
b) To print the report on the printer that you have configured, choose
from the top of the screen.
c) To export the report as a PDF file or to choose a different printer, choose
Print Menu
from the top of the screen.
d) To return to the test result screen, choose
98.
Print
from the bottom right of the
Return
from the menu bar.
Print
93
76/171