Hygiene and cleaning
8.2

Cleaning instructions for EasyOne Pro/LAB

Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
Some parts of EasyOne Pro/LAB are sensitive to certain disinfectants or
fluids. To prevent damage to EasyOne Pro/LAB, follow these cleaning
instructions.
CAUTION
ELECTRIC SHOCK, FAILURE OF EASYONE PRO/LAB, AND FALSE DIAGNOSIS
Make sure that no fluid penetrates the inside of the flow sensor or of
EasyOne Pro/LAB while cleaning.
To clean the flow sensor, do not immerse the flow sensor in any fluid.
1 Wipe down the device surface with a cloth moistened with alcohol
(maximum 70%), avoiding the display, the rear panel, and the temperature
and humidity sensor.
2 To clean the rear panel and the temperature and humidity sensor, use a dry
cloth.
3 To clean the display, use a soft, dry cloth.
 In case of severe contamination, you can use a cloth moistened with
alcohol (maximum 50%).
4 Wipe down the outside of the flow sensor (including the cable), the outside
of the motor & valve unit (including the outside of the one-way valve), and
the outside of the patient tube with a cloth moistened with alcohol
(maximum 70%).
5 Replace the gas supply tube each year.
6 To clean the gas tube, use a cloth moistened with alcohol (maximum 70%).