Calibration check
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
4 Attach the nozzle of the spirette cal check adapter to the spirette and make
sure that the piston is fully inserted.
The spirette snaps into the nozzle of the spirette cal check adapter.
5 Choose
Utilities > Check Calibration > Syringe Calibration Check
confirm the message that is displayed.
6 Wait for the baseline to be set.
7 Continuously push and pull the syringe repeatedly until a green quality
message is displayed.
Try to reproduce the same flow-volume curve with each repetition.
8 Repeat step 7 until the green Accuracy confirmed message is displayed.
One test with three trials is required for the Single Flow calibration
check. For the Multi Flow calibration check, three tests comprising three
trials, each with different flow rates, are required.
Calibration checks are filed in the history menu as patients with the last
name Calibration Check.
List of troubleshooting solutions, 127
Contact information, 4
and
126
123/171