Breathing maneuvers for all available tests
7.4

Performing the breathing maneuver for the MVV test

Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
With the maximum voluntary ventilation test, the patient fully inhales and
fully exhales repeatedly.
Required material
A nose clip
Prerequisite
The temperature and humidity sensor measures the required data
automatically. However, check these values for plausibility. For example,
if the temperature and humidity sensor is too close to the fan on the back
of EasyOne Pro/LAB, the measured temperature and humidity can be
wrong.
If you choose to adjust the values indicated by the temperature and
humidity sensor, you must enter the environment temperature within 1°C
or 1.8°F accuracy for this test.
1 Tell the patient to breathe at rest.
2 Tell the patient to fully exhale.
3 Tell the patient to take the spirette into the mouth and to seal the lips
around the spirette.
 The patient must not block the opening with the tongue or teeth or bite
down excessively on the spirette.
4 Tell the patient to fully inhale and fully exhale repeatedly for an
uninterrupted period of at least 12 seconds.
5 Tell the patient to take the spirette out of the mouth and to breathe
normally again.
 Performing lung function tests, 48