Performing lung function tests
6
Performing lung function tests
6.1
Overview of the spirometry workflow
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
6.2
9.1
6.5.1
Yes
Add trial, add post,
or create new test
Repeat trial
No
General workflow for basic spirometry without gas supply
CAUTION
ELECTRIC SHOCK, PATIENT HEALTH HAZARD, AND FALSE DIAGNOSIS
Damage to casing or broken components and/or disposables may expose the
patient or technician to dangerous voltage. Small parts can be ingested or in-
haled by the patient. Measurement errors can cause false diagnosis.
Check the device for visible damage before on-site installation and before us-
age.
Performing spirometry
Instruct the patient
Optional: Add new patient
Prepare spirette
Select patient
Choose a test
Patient has performed test
within previous 24 h?
No
Set baseline
Patient performs breathing maneuver
Quality message ok?
Yes
3 successful trials completed?
Yes
End test
Dispose of spirette
Print report
Add trial
No
93
48/171