Performing lung function tests
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
2 CAUTION! An incorrectly inserted spirette can lead to inaccurate
measurement, false results, and false diagnosis.
Fully insert the spirette into the spirette holder as depicted below, but keep
the partly unwrapped wrapper on the mouthpiece of the spirette.
The shape of the spirette and the spirette holder guide the orientation
of the spirette. You can only insert the spirette fully if it is positioned
correctly.
3 On the main menu of EasyOne Pro/LAB, choose Patients.
4 Select the patient that you want to test.
5 At the bottom of the screen, choose Test.
6 Choose Provocation.
The provocation protocol screen is displayed.
7 Choose Start.
8 If the environment conditions screen is displayed, verify or adjust the
current values and choose
The temperature and humidity sensor measures the required data
automatically. However, check these values for plausibility. For example, if
the temperature and humidity sensor is too close to the fan on the back of
EasyOne Pro/LAB, the measured temperature and humidity can be wrong.
If you choose to adjust the values indicated by the temperature and
humidity sensor, you must enter the environment temperature within 1°C
or 1.8°F accuracy for the FVL test.
Confirm
>>.
93
61/171