Servicing and troubleshooting
12.3
About reactivating EasyOne Pro/LAB after storage
12.4
Exporting logging information
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
There is no specific procedure for reactivating EasyOne Pro/LAB after storage.
Connect the components and check EasyOne Pro/LAB for correct operation,
according to the topics that are referenced below.
For support, contact your EasyOne dealer or the ndd Servicing Department.
Connecting the components, 21
Checking for correct operation of EasyOne Pro/LAB, 127
Contact information, 4
EasyOne Pro/LAB can export logging information. The ndd Servicing
Department or your own technical staff may require logging information to
determine the source of a problem.
1 On the back of EasyOne Pro/LAB, connect the USB flash drive to a USB port.
2 Choose
Utilities >
Advanced.
You are prompted for the default password.
3 Enter the default password.
The default password is: 8005
4 Choose Export.
5 Select
Logging Information
6 Choose Export.
7 Navigate to and select the USB flash drive.
8 Choose OK.
A confirmation is displayed when the export has been successful.
Contact information, 4
and deselect all other check boxes.