Introduction
3.2.3
Connecting the components
A
B
A
Connector for the temperature and humidity sensor
C
Flow sensor holder
E
Connector for the flow sensor
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
The first-time setup is typically performed by ndd or an ndd distribution
partner. You only need to connect or reconnect components, if you move
EasyOne Pro/LAB or if you exchange components.
CAUTION
ELECTRIC SHOCK, PATIENT HEALTH HAZARD, AND FALSE DIAGNOSIS
Damage to casing or broken components and/or disposables may expose the
patient or technician to dangerous voltage. Small parts can be ingested or in-
haled by the patient. Measurement errors can cause false diagnosis.
Check the device for visible damage before on-site installation and before us-
age.
B
D
1 Plug the power cable into the power port on the back of EasyOne Pro/LAB
and into a power outlet. (B)
2 On the back of EasyOne Pro/LAB, connect the temperature and humidity
sensor. (A)
Make sure that the sensor extends away from EasyOne Pro/LAB because
heat from EasyOne Pro/LAB can affect the measurement of the sensor.
Power port
Connector for the patient tube
C
D
E
35
21/171