Electromagnetic Compatibility (Emc); General; Environment; Emc Conformance - NDD EasyOne Pro Operator's Manual

Hide thumbs Also See for EasyOne Pro:
Table of Contents

Advertisement

Appendix
A.4

Electromagnetic Compatibility (EMC)

A.4.1

General

A.4.1.1

Environment

A.4.1.2

EMC conformance

A.4.2

Safety information

A.4.3

Compliant cables and accessories

A.4.4

Wireless module

Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
See:
 Intended use of the EasyOne Pro Respiratory Analysis System, 3
EasyOne Pro/LAB is EMC-tested in conformity with the requirements of
rd
IEC 60601-1-2:2007 3
th
60601-1-2:2014 4
Edition (according clause 7 and 8.9, tables 4 to 9).
EasyOne Pro/LAB is a medical device that requires special safety precautions
and must be installed and placed in operation in accordance with the
information given in tables below.
See:
 Safety information regarding electromagnetic compatibility, 12
WARNING
INCREASED EMISSIONS OR DECREASED IMMUNITY
The use of accessories, transducers and cables other than those specified
may result in increased emission or decreased immunity performance of the
equipment or system.
The table below lists cables, transducers, and other applicable accessories
with which the manufacturer claims EMC compliance.
Accessories that do not affect EMC compliance are not included in the
following list.
Ethernet cable, 1.5 m, shielded
VGA cable including 2 ferrites, 1.8 m, shielded
Flow sensor cable, 1.7 m, shielded
Temperature / humidity sensor cable, 0.4 m, shielded
 Accessories with which the manufacturer claims EMC compliance
No wireless module is included.
edition (see the following tables) and IEC