12.3

Ethnic Correction

While some predicted normal studies take into account the differences between certain ethnic groups,
most studies used for spirometry were conducted on Caucasian subjects, and are therefore most ap-
propriate for use with Caucasian patients. When entering patient information, you are presented with a
list of ethnic options. In the system configuration there are four Ethnic Correction settings that allow
you to customize the amount of adjustment that is made when African-American, Hispanic, Asian, or
other is chosen during patient data entry. The adjustment is made to the Caucasian values.
There is an exception to this function. When specific values are available for the chosen normal and
ethnic group they will be used rather than the correction entered in the configuration. Refer to Section 8
for instructions on setting the Ethnic Correction Configuration.
The American Thoracic Society s publication, Lung Function Testing: Selection of Reference Values and
Interpretative Strategies [8] provide guidance on the subject of ethnic correction. This paper recom-
mends using 88% of the Caucasian values when testing African-American patients, and provides gen-
eral guidance in selecting adjustments for other ethnic groups.

13 Hygiene and Servicing of the Instrument

EasyOne
has been designed to minimize maintenance and servicing effort if the instrument is used
correctly.
When you use the spirette
cleaning, you simply exchange the respiratory tube. In order to ensure absolute hygiene, we recom-
mend that the spirette
Caution:
Always exchange the spirette
way of absolutely preventing transmission of diseases.
Use a damp cloth to clean the spirometer housing and the base unit. Use a soft cloth and alcohol (e.g.
isopropyl alcohol) for particularly thorough cleaning.
Caution:
Avoid fluid penetrating the spirette
cleaning the spirometer.
Please follow the instructions for changing batteries: Open the battery cover on the backside of the in-
strument. Remove the empty batteries. Please insert two new Alkaline batteries (Type AA, 1,5V) into the
battery case and close the battery cover.
Please consult your EasyOne
the event of defects or malfunctions.
Proceed as follows to check correct operation of your instrument:
1. Check calibration. Please read Chapter 14.
2. Conduct a spirometry test on yourself.
3. Ensure that the results are plausible and that you can print out the report as required.
Consult your EasyOne
respiratory tube, you do not need to clean the instrument. Instead of
be used only once.
dealer or call the ndd Servicing Dept., number +41 (44) 445 29 70 in
dealer if you encounter problems with one of these points.