On-Site Operation; Remote Operation - Endress+Hauser Prosonic M FMU40 Technical Information

Ultrasonic level measurement
Hide thumbs Also See for Prosonic M FMU40:
Table of Contents

Advertisement

On-site operation

Remote operation

28
Operation with VU331
The LC-Display VU331 allows configuration via 3 keys directly at the instrument. All device functions can be
set through a menu system. The menu consists of function groups and functions. Within a function, application
parameters can be read or adjusted. The user is guided through a complete configuration procedure. For ease
operation can choose between 7 language: (de: german; en: english; es: spanish; fr: french; it: italian; ja:
japanese; nl: dutch).
ENDRESS + HAUSER
+
E
Function groups -> Functions
X
HOME
X
F
FG00
FG01
F
O
FG02
FG03
FG04
FG05
S
FG06
FG07
...
Operation with the Field Communicator 375, 475
All device functions can be adjusted via a menu operation with the Field Communicator 375, 475.
Note!
Further information on the handheld unit is given in the respective operating manual included in the transport
bag of the Field Communicator 375, 475.
Operation with FieldCare
FieldCare is Endress+Hauser's FDT based Plant Asset Management Tool. It can configure all intelligent field
devices in your plant and supports you in managing them. By using status information, it also provides a simple
but effective means of checking their health. Hardware and software requirements you can find on the internet:
www.endress.com È select your country È search: FieldCare È FieldCare È Technical Data.
The FieldCare supports the following functions:
• Configuration of transmitters in online operation
• Signal analysis via envelope curve
• Tank linearisation
• Loading and saving of instrument data (Upload/Download)
• Documentation of measuring point
Connection options
• HART with Commubox FXA195 and the USB port on a computer
• PROFIBUS PA via segment coupler and PROFIBUS interface card
• Commubox FXA291 with ToF Adapter FXA291 (USB) via service interface
Symbol
Selection list
F
F000 F001 F002 F003 F004 ...
X
F
O
S
X
Envelope
curve
Headline
Position indicator
Main value
Help text
L00-FMU4xxxx-07-00-00-en-004
Endress+Hauser
Prosonic M
Unit

Advertisement

Table of Contents
loading

Table of Contents