PerkinElmer AAnalyst 700 User Manual page 95

Tomic absorption spectrometer
Table of Contents

Advertisement

7. Unpack the new lamp assembly, taking care to hold it only by the metal mount
to prevent fingermarks on the lamp envelope.
8. Slip the slot at the base of the lamp mount over the stud on the lamp bracket
and secure the lamp assembly with the thumbscrew saved in step 6.
9. Wipe the lamp envelope with a soft cloth moistened with alcohol to remove
dirt, since this would otherwise be burned in when the lamp is hot.
10. Insert the deuterium lamp plug into the socket in the lamp compartment.
The plug is asymmetric and can only be inserted in one direction.
11. Reinstall the lamp assembly:
1. Carefully lower the deuterium lamp assembly into the lamp compartment
2. Screw in the two captive screws on the lamp cover.
12. Close the top cover:
1. Swing down the cover to the closed position.
2. Screw in the two securing screws; one on the left and one on the right.
Since the deuterium lamp is fitted with as prealigned mount, you do not normally
need to align the lamp.
Note: Do not attempt to make adjustments to the lamp bracket assembly. If there
is indication that the alignment is not correct (e.g., low energy for the BG bar
graph in the Align Lamps or Continuous Graphics windows), please contact your
local PerkinElmer office.
0993-5255
until the lamp cover fits in place.
System Maintenance
4-13