EDLs

You must use the lamp holders numbered 1 through 4 for EDLs.
1. Open the lamp compartment cover by lifting it at the front and swinging it
back.
2. If necessary, assemble the lamp by sliding the lamp driver into the lamp
sleeve for the required element. Depress the locking pin as the sleeve slides
into place and make sure that the locking pin extends through the locating
hole in the sleeve.
Locking pin
3. Slide the lamp carefully into the selected lamp holder until it butts up against
the end-stop inside the holder.
4. Insert the coding plug over the receptacle on the lamp holder.
5. Insert the lamp connector into the correct EDL socket at the left-hand side of
the lamp compartment. Use the receptacle that has the same number as the
lamp holder.
6. Close the lamp compartment cover.
Coding plug
EDL socket
0993-5255
Lamp sleeve
Lamp driver
Preparing the System
Coding plug
3
5
7
2
4
6
8
3-9