Preparing the System
8. Switch on the computer.
If your computer has a power management system, switch this off.
9. Install a Lumina hollow cathode lamp in any holder in the lamp compartment
(procedure: page 3-7).
A hollow cathode lamp must be installed in at least one holder before you switch on
the spectrometer.
10. Switch on the spectrometer system with the operational on/off switch located
on the front panel.
11. If you need the diluter, autosampler, or flow injection system, switch on the
respective systems.
12. If you have changed the configuration of the system, such as installed a
different autosampler or added or removed any accessories, use the
Reconfigure
WinLab).
13. Start AA WinLab: In the AA WinLab group, double click on the AA WinLab
icon.
14. Set up the various parts of the system:
1. Install any further lamps that you require for the analyses
2. Set up the burner system
3. Set up the graphite furnace system
Procedures for shutting down the system are described on page 3-12.
In an emergency, you can shut down the system as described on page 3-14.
3-4
utility to reconfigure the software (see the Online Help in AA
(procedure: page 3-7).
(see 'Setting up the system for analyses using the flame technique' in the
Burner System User's Guide).
(see 'Setting up the system for analyses using the furnace technique' in the
Graphite Furnace User's Guide).