Connecting external cooling sources

If you intend to run the spectrometer system without the PerkinElmer cooling
system, you must connect another suitable source of cooling, such as the main
water supply or a laboratory thermostat. If you connect the spectrometer system
to the main water supply, make sure that the water pressure is as given in Table
2-1.
Figure 2-5. Connecting the Main Water Supply
A water hose kit with pressure relief valve is provided with the spectrometer
system which you must install in the water supply line to the spectrometer. This
valve opens if the water pressure to the furnace system exceeds 4 bar. This feature
eliminates the risk of damage to the furnace system through excess water
pressure.
The cooling water must conform with the requirements listed in 'External
cooling' on page 6-16.
0993-5255
Spectrometer Installation
Pressure
relief valve
Drain hose
Flow
indicator
Water In
Water Out
2-19