Spectrometer Installation
Caution
Water Pressure – Leaking hose fittings or burst water hoses
If the water inlet pressure is too high, hose fittings may leak or hoses may even
burst.
 For reliable and safe operation, you must install the pressure relief valve kit
B313-0124 in the water supply line.
The hoses provided in the kit are fitted with automatic connectors that incorporate
stop valves; these valves prevent water from running out when the hoses are dis-
connected. The hoses are marked with the direction of the water flow.
Table 2-1. Pressure for the Cooling Water Supply
Check the water pressure to your laboratory and make sure that the water inlet
pressure is set to a value between the minimum value (P
maximum value (P
Make sure that the outlet pressure is never set to a value exceeding the
maximum pressure (P
1. The PerkinElmer service engineer will connect the pressure relief valve to the
water tap (faucet).
2. Make sure that the drain hose from the pressure relief valve hangs down into
a suitable laboratory sink or drain.
3. Connect the inlet hose (marked
pressure relief valve.
4. Connect the other end of the inlet hose to the
In
The connectors are spring loaded. Push them straight onto the fittings until
they click into place.
2-20
Outlet Pressure
P
P
min
max
250
350
kPa
2.5
3.5
bar
36
50
psig
).
max
max
on the spectrometer.
).
) to the fitting at the bottom of the
WATER IN
Furnace Cooling
) and the
min
fitting marked
0993-5255