Signal handling Time average, peak average, peak height, peak area.
Calibration
Application
program
Graphics
Power
requirements
0993-5255
Measurement ranges:
absorbance: –0.5 to 2.000
concentration: 0.00001 to 9999 concentration units
expansion factor: 0.01 to 100
read time: 1 to 60 s
read delay: 0 to 60 s
Automatic zeroing, resloping, and recalibration.
Calibration/recalibration using up to 15 calibration solutions ('standards').
Calibration coefficients determined by a least squares procedure.
Analytical curve technique:
Linear with zero or non-zero intercept.
Non-linear, 2-coefficient, or 3-coefficient, zero or non-zero intercept.
Analyte addition technique:
Zero or non-zero intercept.
Additions calibrate (zero intercept)
Special application program: AA WinLab.
Online help function.
Hard disk or floppy storage of measurement data, signals, and completed
analytical methods.
Special Reporter, Reformatter, and Library Manager applications for
generating data reports and reformatting data for export to other data
handling programs.
High resolution display of the signals.
Simultaneous display of the corrected analyte signal and the background
signal.
Calibration curve display.
Storage and recall of transient absorbance signals.
Hard copy printout of screen graphics on an associated printer.
230 V single-phase, 50/60 Hz, power consumption 5000 VA.
Spectrometer System Description
7-15