PerkinElmer AAnalyst 700 User Manual page 134

Tomic absorption spectrometer
Table of Contents

Advertisement

Spectrometer System Description
Electrical
protection
Safety standards As defined in EN 61010-1-1993 (IEC 1010-1: 1990 + A1 + A2) and
EMC standards As defined in EN 61326-1:1997.
Environmental
requirements
Dimensions
Mass
(weight)
7-16
As defined in EN 61010-1-1993 (IEC 1010-1):
Insulation: Class I
Insulation category (overvoltage category): II
Pollution degree: 2
EN 61010-2-061:1996.
CSA standards: CAN/CSA-C22.2 No. 1010.1-92, Amendment 1 (1992),
2 (1995), and CAN/CSA 61010-2-061 (1995).
UL standards: UL 3101-1 1993 and UL 3111-1 1994.
Compliant to the requirements of Canadian Standards Association (CSA)
and Underwriters Laboratory (UL) as listed in File No. LR 45111.
Certificate of Approval and GS mark from the LGA Bavaria, FRG, Institute
of Equipment Safety Testing and Certification, and CSA-NRTL/C mark
from CSA, Canada, File No. LR 45111.
Ambient temperature: +15 °C to +35 °C
20–80% relative humidity; non-condensing
Width: 1095 mm
Height: 655 mm
Depth: 700 mm
147 kg
Spectrometer alone, not including the computer, autosampler, and cooling
system.
with handles:
1220 mm
with door open:
865 mm
with handles:
770 mm
with autosampler: 1040 mm
0993-5255