System Maintenance

Changing the air filters

To make sure that the electrical components in the spectrometer are properly
cooled, the air filters must allow sufficient air into the instrument. Check the
filters regularly and change them if they are dirty.
You should not need to change the filters more than once every year if the
environmental conditions are as described under Chapter 6, Laboratory
Requirements.
Replacement filters are available; see Chapter 5, Parts and Supplies for the part
number.
1. Carefully pull off the snap-on plastic grid that holds the air filter in place.
2. Replace the filter and push the grid back into place.
4-10
Instrument
Communica tion
230 V
230 V
230 V
50/60 Hz
50/60 Hz
50/60 Hz
F1
F2
F3
4 A (T)
4 A (T)
4 A (T)
230 V
50/60 Hz
Cooling System
Système de Refroidissement
Umlaufkühlu ng
I
0
In In
Out
Out
Entrée
Entrée
Sortie
Sortie
Eingang
Eingang
Ausgang
Ausgang
H O
P
P
P P
Warning
2
min
max
All field wiring must have i ns ulation
250
350
kPa
Furnace Cooling
s uit able for at leas t 300 V
2.5
3.5
bar
Ave rtissement
Refroidissement du Four
36
50
psig
Le cablage à pied d'ouv re doit être
c ertif iés pour un minimum de 300 V
Ofenkühlung
Furnace Gases
Gaz du Four
Ofengase
P
P
min
max
350
400
kPa
3.5
4.0
bar
Ar / N
50.0
58.0
psig
2
Normal Gas
Special Gas
F4
Gaz Norm al
Gaz Spécial
4 A (T)
Normalgas
Spezialgas
Burner Gases
Gaz du Bruleur
Brennergase
P
P
P
P
P
P
min
max
min
max
min
max
450
500
kPa
450
500
kPa
90
100
kPa
4.5
5.0
bar
4.5
5.0
bar
0.9
1.0
bar
65.0
72.5
psig
65.0
72.5
psig
13.0
14.5
psig
Air / Luft
N O
C H
2
2 2
SA
LR 45111
***
Air filters
0993-5255

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for PerkinElmer AAnalyst 700

Table of Contents