Replacing the deuterium lamp

The deuterium lamp for the background corrector has a limited lifetime. After a
prolonged period of operation the lamp will exhibit decreasing radiant intensity
and will ultimately fail.
The lifetime of the deuterium lamp will be prolonged if you do not switch it on
and off frequently.
You can replace the deuterium lamp assembly yourself. The replacement lamp is
fitted in a prealigned mount.
Deuterium
lamp cover
Figure 4-2. Location of Deuterium Lamp inside the Spectrometer
Warning: Electrical Hazard
 To prevent potential injury to yourself and damage to the instrument, switch
OFF all instruments in the system and disconnect them from the line power
supply before you perform the procedures described below.
1. Switch off the spectrometer at the operational on/off switch on the front panel.
2. Switch the circuit breaker on the rear panel of the spectrometer to the off
position (marked ).
3. Disconnect the spectrometer from the line power supply.
0993-5255
System Maintenance
W2.3
4-11