Gas supply requirements

The permanent installation of gas supplies is the responsibility of the user and
should conform to local safety and building regulations.
PerkinElmer recommends that cylinders be stored outside the laboratory and that
gases be piped to the spectrometer system in approved gas lines.
The user must provide the gas supplies, regulators, connectors, and valves.
A double-stage or heated regulator is recommended for nitrous oxide.
Either the gas supplier or PerkinElmer can supply the correct regulators and
valves.
You must be able to reach the on/off valves easily and see the pressure
indicators.
Consult Chapter 1, Safety and Regulatory Information for recommended
safety procedures.
Warning: Compressed Gases
High pressure gas cylinders can be dangerous if mishandled or misused.
 Always handle gas cylinders with caution and observe your local regulations
for the safe handling of gas cylinders.
0993-5255
Laboratory Requirements
6-9