Required Pressures For The Burner Gases - PerkinElmer AAnalyst 700 User Manual

Tomic absorption spectrometer
Table of Contents

Advertisement

Table 3-1. Required Pressures for the Burner Gases
Gas
Acetylene
(C
2
Air
Nitrous oxide
(N
2
Set the outlet gauge pressure for each gas to a value just below the maximum
value (P
Always set the same outlet gauge pressure for air and nitrous oxide.
Never set the outlet gauge pressure to a value exceeding the maximum
pressure (P
Always turn on the air supply since air pressure is required for the
spectrometer's pneumatic system.
Warning: Acetylene – Explosion Hazard
Acetylene can decompose explosively at pressures higher than 103 kPa/
1.03 bar/15 psig.
 Always make sure that the acetylene outlet gauge pressure is below this
value.
0993-5255
kPa
P
min
90
H
)
2
450
450
O)
).
max
).
max
Outlet Gauge Pressure
bar
P
P
P
max
min
max
100
0.9
1.0
500
4.5
5.0
500
4.5
5.0
Preparing the System
psig
P
P
min
max
13.0
14.5
65.0
72.5
65.0
72.5
W5.7
3-5