To Connect Both Argon And A Special Gas - PerkinElmer AAnalyst 700 User Manual

Tomic absorption spectrometer
Table of Contents

Advertisement

To connect both argon and a special gas

Refer to Figure 2-3 on the following page when you connect the furnace gases.
1. Make sure that the gas supplies are turned off.
2. Use the special tube cutter to cut the special gas tube into two pieces for the
two gas supplies. The ends of the tubes must be cut squarely.
3. Connect the gas tubes to the argon (or nitrogen) and special gas pressure
regulators. Use the R ¼'' screw connectors provided or use a connector or
adapter suitable for the gas regulator. Make sure that all the connections are
secure.
4. Pass the gas tubes to the rear of the spectrometer. Make sure that they cannot
be damaged or strained.
5. Push the free ends of the gas supply tubes into the
Gas
Push the tubes into the connectors until the tubes butt up to the end of the
connector.
Note: To disconnect the gas tube from a connector:
Carefully push the black locking ring of the connector into the connector and
pull the tube out of the connector.
0993-5255
connectors on the manifold as shown.
Spectrometer Installation
and
Normal Gas
Special
2-13

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for PerkinElmer AAnalyst 700

Table of Contents