Spectrometer System Description

The gas and cooling connectors

Item Designation
Furnace Cooling
1
2
Furnace Gases
3
4
Burner Gases
5
6
7
7-12
Function
Fitting for connection of the cooling water inlet hose
In
to the spectrometer.
Out
Fitting for connection of the cooling water outlet hose
from the spectrometer.
Ar/N
Connector for the connection of the normal gas supply
2
to the furnace (argon or nitrogen).
Normal Gas
Special Gas
Connector for the connection of the special gas supply
to the furnace. If you are not using a special gas, you
must also connect the normal gas to this connector.
Air/Luft
Fitting for connection of the air supply to the
spectrometer. The air supply is required for the burner
and also for the pneumatic system in the spectrometer.
N
O
Fitting for connection of the nitrous oxide supply to
2
the burner.
C
H
Fitting for connection of the acetylene supply to the
2
2
burner.