14
:DVM Commands
:DVM:ARANge
Command Syntax
:DVM:ARANge <setting>
<setting> ::= {{OFF | 0} | {ON | 1}}
If the selected digital voltmeter (DVM) source channel is not used in
oscilloscope triggering, the :DVM:ARANge command turns the digital
voltmeter's Auto Range capability on or off.
• When on, the DVM channel's vertical scale, vertical (ground level)
• When off, you can adjust the channel's vertical scale and position
Query Syntax
:DVM:ARANge?
The :DVM:ARANge? query returns a flag indicating whether the digital
voltmeter's Auto Range capability is on or off.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also
•
•
•
264
(see
page
790)
position, and trigger (threshold voltage) level (used for the counter
frequency measurement) are automatically adjusted.
The Auto Range capability overrides attempted adjustments of the
channel's vertical scale and position.
normally.
":DVM:SOURce"
on page 269
":DVM:ENABle"
on page 266
":DVM:MODE"
on page 268
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide