:MTESt:RMODe:FACTion:MEASure
Command Syntax
:MTESt:RMODe:FACTion:MEASure <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:RMODe:FACTion:MEASure command sets measuring only mask
failures on or off.
When ON, measurements and measurement statistics run only on
waveforms that contain a mask violation; passing waveforms do not affect
measurements and measurement statistics.
This mode is not available when the acquisition mode is set to Averaging.
Query Syntax
:MTESt:RMODe:FACTion:MEASure?
The :MTESt:RMODe:FACTion:MEASure? query returns the current mask
failure measure setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also
•
•
•
•
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
(see
page
790)
"Introduction to :MTESt Commands"
":MTESt:RMODe:FACTion:PRINt"
":MTESt:RMODe:FACTion:SAVE"
":MTESt:RMODe:FACTion:STOP"
on page 379
on page 398
on page 399
on page 400
21
:MTESt Commands
397