Measure:dutycycle - Agilent Technologies InfiniiVision 2000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 2000 X-Series:
Table of Contents

Advertisement

:MEASure:DUTYcycle

Command Syntax
:MEASure:DUTYcycle [<source>]
<source> ::= {<digital channels> | CHANnel<n> | FUNCtion | MATH
<digital channels> ::= DIGital<d> for the MSO models
<n> ::= 1 to (# of analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :MEASure:DUTYcycle command installs a screen measurement and
starts a duty cycle measurement on the current :MEASure:SOURce. If the
optional source parameter is specified, the current source is modified.
The signal must be displayed to make the measurement. This command is not available if
N O T E
the source is FFT (Fast Fourier Transform).
Query Syntax
:MEASure:DUTYcycle? [<source>]
The :MEASure:DUTYcycle? query measures and outputs the duty cycle of
the signal specified by the :MEASure:SOURce command. The value
returned for the duty cycle is the ratio of the positive pulse width to the
period. The positive pulse width and the period of the specified signal are
measured, then the duty cycle is calculated with the following formula:
Return Format
<value><NL>
<value> ::= ratio of positive pulse width to period in NR3 format
See Also
Example Code
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
(see
page
790)
| WMEMory<r>}
duty cycle = (+pulse width/period)*100
"Introduction to :MEASure Commands"
":MEASure:PERiod"
on page 355
":MEASure:PWIDth"
on page 358
":MEASure:SOURce"
"Example Code"
on page 362
on page 340
on page 361
20
:MEASure Commands
349

Advertisement

Table of Contents
loading

Table of Contents