Measure:risetime - Agilent Technologies InfiniiVision 2000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 2000 X-Series:
Table of Contents

Advertisement

:MEASure:RISetime

Command Syntax
:MEASure: RISetime [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:RISetime command installs a screen measurement and
starts a rise- time measurement. If the optional source parameter is
specified, the current source is modified.
This command is not available if the source is FFT (Fast Fourier Transform).
N O T E
Query Syntax
:MEASure: RISetime? [<source>]
The :MEASure:RISetime? query measures and outputs the rise time of the
displayed rising (positive- going) edge closest to the trigger reference. For
maximum measurement accuracy, set the sweep speed as fast as possible
while leaving the leading edge of the waveform on the display. The rise
time is determined by measuring the time at the lower threshold of the
rising edge and the time at the upper threshold of the rising edge, then
calculating the rise time with the following formula:
Return Format
<value><NL>
<value> ::= rise time in seconds in NR3 format
See Also
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
(see
page
790)
rise time = time at upper threshold - time at lower threshold
"Introduction to :MEASure Commands"
":MEASure:SOURce"
":MEASure:FALLtime"
on page 340
on page 361
on page 350
20
:MEASure Commands
359

Advertisement

Table of Contents
loading

Table of Contents