:SBUS<n> Commands
:SBUS<n>:LIN:SAMPlepoint
Command Syntax
:SBUS<n>:LIN:SAMPlepoint <value>
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
The :SBUS<n>:LIN:SAMPlepoint command sets the point during the bit
time where the bit level is sampled to determine whether the bit is
dominant or recessive. The sample point represents the percentage of time
between the beginning of the bit time to the end of the bit time.
The sample point values are not limited by the baud rate.
N O T E
Query Syntax
:SBUS<n>:LIN:SAMPlepoint?
The :SBUS<n>:LIN:SAMPlepoint? query returns the current LIN sample
point setting.
Return Format
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
See Also
482
(see
page
790)
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 598
":SBUS<n>:LIN:TRIGger"
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
on page 589
on page 487