Table 37 :TRIGger[:EDGE] Commands Summary (continued)
Command
:TRIGger[:EDGE]:SLOPe
<polarity> (see
page
605)
:TRIGger[:EDGE]:SOURc
e <source> (see
page
606)
Table 38 :TRIGger:GLITch Commands Summary
Command
:TRIGger:GLITch:GREat
erthan
<greater_than_time>[s
uffix] (see
page
:TRIGger:GLITch:LESSt
han
<less_than_time>[suff
ix] (see
page
610)
:TRIGger:GLITch:LEVel
<level> [<source>]
(see
page
611)
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
Query
:TRIGger[:EDGE]:SLOPe
? (see
page
:TRIGger[:EDGE]:SOURc
e? (see
page
Query
:TRIGger:GLITch:GREat
erthan? (see
page
609)
609)
:TRIGger:GLITch:LESSt
han? (see
page
:TRIGger:GLITch:LEVel
? (see
page
Options and Query Returns
<polarity> ::= {POSitive |
605)
NEGative | EITHer | ALTernate}
<source> ::= {CHANnel<n> |
606)
EXTernal | LINE | WGEN} for the
DSO models
<source> ::= {CHANnel<n> |
DIGital<d> | EXTernal | LINE |
WGEN} for the MSO models
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
Options and Query Returns
<greater_than_time> ::=
floating-point number in NR3
format
[suffix] ::= {s | ms | us | ns |
ps}
<less_than_time> ::=
610)
floating-point number in NR3
format
[suffix] ::= {s | ms | us | ns |
ps}
For internal triggers, <level>
611)
::= .75 x full-scale voltage from
center screen in NR3 format.
For external triggers (DSO
models), <level> ::= ±(external
range setting) in NR3 format.
For digital channels (MSO
models), <level> ::= ±8 V.
<source> ::= {CHANnel<n> |
EXTernal} for DSO models
<source> ::= {CHANnel<n> |
DIGital<d>} for MSO models
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format