Power Supply

Power Monitoring

User Interface Printed Circuit Board

NPB-40
The power supply provides operating voltages to the SpO
interface PCB. These voltages are supplied to the:
SpO
PCB power monitoring function
2
SpO
PCB circuits
2
user interface PCB power supply
user interface PCB CPU
user interface PCB audio circuits
The SpO
power monitoring function monitors and controls the outputs of the
2
SpO
power supply. The SpO
2
interface CPU and SpO
2
supply output signals.
The user interface PCB provides:
User setup of the NPB-40 via the keypad
Patient %SpO2 and pulse rate via the display
NPB-40 status via the display and audio function
Printed reports via the IrDA (Infrared Data Association) and external
printer
Display readability in dark environments via the display backlight
Operating voltages for the user interface circuits
Operating voltage monitoring
power monitoring function provides the user
2
patient interface with status information on the power
Technical Discussion
PCB and the user
2
79