Other Messages

Adjust Contrast Up,
Down
Clock Settings Lost
Data In Sensor
Data Type: Event/
SpO
2
Data Type: Event/
SpO
+BPM
2
Defaults Lost
Delete Trend?
Invalid Blip Vol
N-595

Table 10: Error Code Problems

Error
Meaning
Code
4
Dead battery
80
Institutional parameter lost
81
Current settings lost
82
Clock setting lost
In addition to the error codes listed in Table 10 on page 49, the following
messages may be encountered:
When the user attempts to adjust the contrast of the display by pressing or
pressing and holding the CONTRAST button.
If the N-595 detects that the real time clock has stopped running. This will
usually occur when both battery and AC power are lost.
When a sensor containing data is connected to the N-595.
When a blank sensor is connected to a monitor with Data Type set to SpO
When a blank sensor is connected to a monitor with Data Type set to
SpO
+ BPM.
2
If the N-595 detects that the power-on settings have been lost.
When the user attempts to delete trend data from memory by pressing the
DELETE softkey.
When the user attempts to save current settings as power-on defaults and the
blip volume is 0.
Troubleshooting
Possible Solution or Problem
Identification
Refer to Table 5 on page 44.
User interface module: Institutional
EEPROM section CRC corrupt
User interface module: current
EEPROM section CRC corrupt
Re-enter the date and time.
.
2
49