Other Messages; Table 8: Error Codes And Messages - Nellcor N-395 Operator's Manual

Table of Contents

Advertisement

Table 8: Error Codes and Messages
Other Messages
Error
Code
4
80
81
82
I11 addition to the messages listed in Table 8, the following
messages may be encouiltered:
SENSOR DISCONNECTED
-
The seilsor has disconnected from
the cable, the cable has disconnected from the monitor, or the
sensorlcable wiring is defective. Press the ALARM SILENCE
Butt011 to silence the alarm. Check the connections. If this does
not correct the problem, replace the sensor andlor cable.
Error
Message
LOW
BATTERY
DEFAULTS
LOST
SETTINGS
LOST
CLOCK
SETTING
LOST
Action
The battery is discharged to a critically low level.
Check to ensure that the voltage selector switch on the
rear panel is set to the proper voltage.
Turn the monitor off and let it charge for
approximately 10 minutes and then reattempt to
turn
the unit on. If the error code is still present, turn the
unit off and let it continue to charge. If the monitor
has been charged for 30 minutes and the error code is
still present, notify service personnel.
The current power-on default settings have been lost
and returned to factory defaults. Qualified service
personnel can use the service manual to restore the
desired power-on default settings.
The current settings (for example, alarm limits, alarm
and pulse beep volumes, alarm silence duration) have
been lost and returned to power-on defaults. Turn the
monitor off and Gack on again. If it is necessary to
have settings different from the power-on default
settings, turn the monitor off and back on again, and
reenter the desired settings.
The date and time settings have been lost. Re-enter the
date and time.