MAINTENANCE

Service

Cleaning

Battery Testing

Determining Software Version
Technical Assistance
Returning the N-200
SERVICE
WARNING: For continued protection against fire,
replace fuses only with those of the same type and
rating.
The N-200 requires no routine service other than periodic
battery replacement and any performance testing mandated by
the operator's institution. The Troubleshooting section
discusses potential difficulties, their possible causes, and
suggestions for resolving them. Complete service instructions,
including performance tests checks, are contained in the
NELLCOR N-200 service manual.
CLEANING
Caution: Do not immerse the N-200 in liquid or use
caustic or abrasive cleaners.
To clean the N-200's surfaces, dampen a cloth with a
commercial, nonabrasive cleaner and wipe the top, bottom, and
front surfaces lightly. Do not spray or pour any liquid directly
on the N-200, interface/powerbase, or their accessories. Do not
allow any liquid to come in contact with the power connector,
fuse holder, or switches. Do not allow any liquid to penetrate
the connectors or openings in the chassis.
BATTERY TESTING
The N-200 is equipped with sealed lead-acid batteries. Battery
capacity decreases over time due to several factors, such as
temperature, depth of discharge, and number of charge/
discharge cycles.