Transducer Block 1 Object; Transducer Block - Emerson Micro Motion 2700 Configuration And Use Manual

With profibus-pa
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Model 2700 PROFIBUS Block Parameters
Table F-5
Transducer block 1 parameters (continued)
Index
Parameter Mnemonic
238
UI_ProcessVariables
(LDO_VAR_7_CODE)
(DD Name: Display Variable 7)
239
UI_ProcessVariables
(LDO_VAR_8_CODE)
(DD Name: Display Variable 8)
240
UI_ProcessVariables
(LDO_VAR_9_CODE)
(DD Name: Display Variable 9)
241
UI_ProcessVariables
(LDO_VAR_10_CODE)
(DD Name: Display Variable
10)
242
UI_ProcessVariables
(LDO_VAR_11_CODE)
(DD Name: Display Variable
11)
243
UI_ProcessVariables
(LDO_VAR_12_CODE)
(DD Name: Display Variable
12)
244
UI_ProcessVariables
(LDO_VAR_13_CODE)
(DD Name: Display Variable
13)
245
UI_ProcessVariables
(LDO_VAR_14_CODE)
(DD Name: Display Variable
14)
246
UI_ProcessVariables
(LDO_VAR_15_CODE)
(DD Name: Display Variable
15)
247
UI_UpdatePeriodmsec
248
EMPTY
249
UI_Language
250
STATUS_LED_TEST
251
EMPTY
252
EMPTY
253
EMPTY
254
Transducer Block1 VIEW1
F.4.1

Transducer block 1 object

Table F-6 shows the transducer block 1 object.
170
Definition
Display the Variable#7
associated with the code on the
LDO
Display the Variable#8
associated with the code on the
LDO
Display the Variable#9
associated with the code on the
LDO
Display the Variable#10
associated with the code on the
LDO
Display the Variable#11
associated with the code on the
LDO
Display the Variable#12
associated with the code on the
LDO
Display the Variable#13
associated with the code on the
LDO
Display the Variable#14
associated with the code on the
LDO
Display the Variable#15
associated with the code on the
LDO
Display update period
(milliseconds)
Display language selection
Simulate the Status LED
Message
Data Type/
Size
Type
Structure
ENUM
Unsigned16
2
ENUM
Unsigned16
2
ENUM
Unsigned16
2
ENUM
Unsigned16
2
ENUM
Unsigned16
2
ENUM
Unsigned16
2
ENUM
Unsigned16
2
ENUM
Unsigned16
2
ENUM
Unsigned16
2
VARIABLE
Unsigned16
2
ENUM
Unsigned16
2
ENUM
Unsigned
2
16
Store
Default
Access
Enumerated List of
/Rate
Value
Values / Range
(HZ)
S
251
R/W
Same as
LDO_VAR_2_CODE
S
251
R/W
Same as
LDO_VAR_2_CODE
S
251
R/W
Same as
LDO_VAR_2_CODE
S
251
R/W
Same as
LDO_VAR_2_CODE
S
251
R/W
Same as
LDO_VAR_2_CODE
S
251
R/W
Same as
LDO_VAR_2_CODE
S
251
R/W
Same as
LDO_VAR_2_CODE
S
251
R/W
Same as
LDO_VAR_2_CODE
S
251
R/W
Same as
LDO_VAR_2_CODE
S
200m
R/W
Between 100 to
s
10,000
S
English R/W
0 = English
1 = German
2 = French
3 = Not used
4 = Spanish
S
0x000
R/W
0=off, 1=green,
0
2=red, 3=yellow,
4=flashing, add 4 to
flash
Model 2700 Transmitter with PROFIBUS-PA
Modbus
Register /
Coil
R-1123
R-1124
R-1125
R-1126
R-1127
R-1128
R-1129
R-1130
R-1131
R-2621
R-1359
R 5006

Advertisement

Table of Contents
loading

Table of Contents