Startup
2.3

Setting the node address

The factory default setting for the node address is 126. To set the node address:
With the display, choose
With ProLink II, choose
With a PROFIBUS host, use the change address function of the host.
2.4

Configuring the analog input function block channels

You can set each of the transmitter's AI function blocks to measure one transducer block channel. The
AI blocks are set at the factory to a default setting that is adequate for most applications, but you can
change the assignment of the AI blocks to meet special needs.
The default channel configuration for each block is shown in Table 2-1.
Table 2-1
Default channel configuration
Block
AI 1
AI 2
AI 3
AI 4
The available transducer block channels are shown in Table 2-2.
Table 2-2
Process variables by transducer block channel
Slot
11 (0x0B)
11 (0x0B)
11 (0x0B)
11 (0x0B)
11 (0x0B)
11 (0x0B)
11 (0x0B)
12 (0x0C)
12 (0x0C)
12 (0x0C)
12 (0x0C)
12 (0x0C)
12 (0x0C)
12 (0x0C)
12 (0x0C)
12 (0x0C)
8
OFF-LINE MAINT > CONFG > ADDRESS PBUS
ProLink > Configuration > Device (Profibus) > Profibus Address
Default channel
Mass flow
Temperature
Density
Volume flow
Channel value
Index
Value
17 (0x11)
0x0B11
21 (0x15)
0x0B15
25 (0x19)
0x0B19
29 (0x1D)
0x0B1D
64 (0x40)
0x0B40
114 (0x72)
0x0B72
160 (0xA0)
0x0BA0
29 (0x1D)
0x0C1D
30 (0x1E)
0x0C1E
31 (0x1F)
0x0C1F
32 (0x20)
0x0C20
33 (0x21)
0x0C21
47 (0x2F)
0x0C2F
48 (0x30)
0x0C30
49 (0x31)
0x0C31
50 (0x32)
0x0C32
Default units
kg/s
K
kg/l
3
m
/h
Process variable
Volume flow
Mass flow
Density
Temperature
Gas standard volume flow
Pressure
Drive gain
Petroluem measurement – corrected density
Petroluem measurement – corrected volume flow
Petroluem measurement – average corrected density
Petroluem measurement – average corrected temp
Petroluem measurement – CTL
Concentration measurement – reference density
Concentration measurement – specific gravity
Concentration measurement – standard volume flow
Concentration measurement – net mass flow
Model 2700 Transmitter with PROFIBUS-PA
.
.