Calibration
Figure 3-17
D3 or D3-and-D4 density calibration – ProLink II
Close shutoff valve
downstream from sensor
3.8
Performing temperature calibration
Temperature calibration is a two-point procedure: temperature offset calibration and temperature
slope calibration. The entire procedure must be completed without interruption.
To perform a temperature calibration, you must use ProLink II. Refer to the menu flowchart in
Figure 3-18.
44
D3 Calibration
Fill sensor with D3 fluid
ProLink Menu >
Calibration >
Density cal – Point 3
Enter density of D3 fluid
Do Cal
Calibration in Progress light
turns red
Calibration in Progress light
turns green
Close
D4 Calibration
Fill sensor with D4 fluid
ProLink Menu >
Calibration >
Density cal – Point 4
Enter density of D4 fluid
Do Cal
Calibration in Progress light
turns red
Calibration in Progress light
turns green
Close
Done
Done
Model 2700 Transmitter with PROFIBUS-PA