Troubleshooting
Table 6-4
Status alarms and remedies (continued)
Display
code
Description
A028
Sensor/transmitter write failure
A030
Hardware/software incompatible
A031
Low Power
A032
Smart Meter Verification in progress
and outputs fixed
A033
Sensor OK / Tubes Stopped by
Process
A034
Smart Meter Verification failed
A035
Smart Meter Verification aborted
A102
Drive overrange
A103
Data loss possible
A104
Calibration in progress
A105
Slug flow
A107
Power reset occurred
A116
API temperature outside standard
range
A117
API density out of limits
A120
Concentration measurement:
unable to fit curve data
A121
Concentration measurement:
extrapolation alarm
A131
Smart Meter Verification in progress Allow the procedure to complete.
A132
Simulation mode active
Configuration and Use Manual
Possible remedies
Cycle power to the meter.
The flowmeter might need service. Contact Micro Motion Customer
Service.
The loaded software is not compatible with the programmed board
type. Contact Micro Motion.
The core processor is not receiving enough power. Check the power
supply to the transmitter, and check power wiring between the
transmitter and the core processor (4-wire remote installations only).
Allow the procedure to complete.
If desired, abort the procedure and restart with outputs set to
Continue Measurement.
No signal from LPO or RPO, suggesting that sensor tubes are not
vibrating. Verify process. Check for air in the flow tubes, tubes not
filled, foreign material in tubes, or coating in tubes.
Rerun the test. If the test fails again, see Section 3.4.3.
If desired, read the abort code. See Section 3.4.3, and perform the
appropriate action.
Excessive or erratic drive gain. See Section 6.12.3.
Check the sensor coils. See Section 6.14.
Cycle power to the transmitter.
The transmitter might need service. Contact Micro Motion Customer
Service.
Allow the flowmeter to complete calibration.
Allow slug flow to clear from the process.
See Section 6.10.
No action is necessary.
Verify process.
Verify API reference table and temperature
configuration. See Section 4.6.
Verify process.
Verify API reference table and temperature
configuration. See Section 4.6.
Verify enhanced density configuration.
Verify process temperature.
Verify process density.
Verify enhanced density configuration.
If desired, abort the procedure and restart with outputs set to Fault.
Disable sensor simulation mode. See Section 5.5.