Emerson Micro Motion 2700 Configuration And Use Manual page 169

With profibus-pa
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Model 2700 PROFIBUS Block Parameters
Table F-5
Transducer block 1 parameters (continued)
Index
Parameter Mnemonic
143
PA_StatusWords5
(DD Name: Alarm Five Status)
144
PA_StatusWords6
(DD Name: Alarm Six Status)
145
PA_StatusWords7
(DD Name: Alarm Seven
Status)
146
PA_StatusWords8
(DD Name: Alarm Eight Status)
147
SYS_DigCommFaultAction
Code
(DD Name Digital Comm Fault
Action)
148
DB_SYS_TimeoutValueLMV
(DD Name : Last Measured
Value Timeout)
Configuration and Use Manual
Definition
Message
Type
Status Word 5
ENUM
Status Word 6
ENUM
Status Word 7
ENUM
Status Word 8
ENUM
Fault Limit Code
ENUM
Last Measure Value Fault
VARIABLE
Timeout
Data Type/
Size
Store
Default
Access
Structure
/Rate
Value
(HZ)
BIT_ENUM
2
D/20
-
R
ERATED
BIT_ENUM
2
D/20
-
R
ERATED
BIT_ENUM
2
D/20
---
R
ERATED
BIT_ENUM
2
D/20
---
R
ERATED
Unsigned16
2
S
0
R/W
Unsigned16
2
S
0
R/W
Enumerated List of
Modbus
Values / Range
Register /
Coil
0x0001 = Boot sector
R-423
(CP)
0x0002 = Not Used
0x0004 = Not Used
0x0008 = Not Used
0x0010 = Not Used
0x0020 = Not Used
0x0040 = D3 in
progress
0x0080 = D4 in
progress
0x0100 = Not Used
0x0200 = Not Used
0x0400 = Temp slope
in progress
0x0800 = Temp offset
in progress
0x1000 = FD in
progress 0x2000 =
D2 in progress
0x4000 = D1 in
progress
0x8000 = Zero in
progress
0x0001 = Not Used
R-424
0x0002 = Not Used
0x0004 = Not Used
0x0008 = Not Used
0x0010 = Not Used
0x0020 = Not Used
0x0040 = Not Used
0x0080 = Not Used
0x0100 = DE0 active
0x0200 = DE1 active
0x0400 = DE2 active
0x0800 = DE3 active
0x1000 = DE4 active
0x2000 = Not Used
0x4000 = Not Used
0x8000 = Incorrect
Board Type
(A30)
0x0001 = K1/FCF
R-433
combination
Unrecognized.
0x0002 = Warming
Up
0x0004 = Low Power
(A31)
0x0008 = Tube Not
Full (A33)
0x0010 = Meter
Verification / Outputs
in fault (A32)
0x0020 = Meter
Verification / Outputs
at last value (A131)
0x0040 = PIC UI
EEPROM Error
0x0080 = Not Used
0x0100 = Not Used
0x0200 = Not Used
0x0400 = Not Used
0x0800 = Not Used
0x1000 = Not Used
0x2000 = Not Used
0x4000 = Not Used
0x8000 = Not Used
0x0001 = Not used
R-434
0x0002 = Not Used
0x0004 = Not Used
0x0008 = Not Used
0x0010 = Not Used
0x0020 = Not Used
0x0040 = Not Used
0x0080 = Not Used
0x0100 = Not Used
0x0200 = Not Used
0x0400 = Not Used
0x0800 = Not Used
0x1000 = Not Used
0x2000 = Not Used
0x4000 = Not Used
0x8000 = Not Used
0 = Upscale
R-124
1 = Downscale
2 = Zero
3 = NAN
4 = Flow goes to zero
5 = None
0 to 60
R-314