Calibration
Figure 3-9
Zeroing procedure – EDD
Calibration >
Zero Cal
Modify zero time if
Start Zero Cal
Stop flow through
Figure 3-10
Zeroing procedure – Bus parameters
Modify zero time
(if desired)
Initiate zero
Check status
Check for failure
alarms
Check zero value
38
desired
Execute
sensor
OK
Block: Transducer Block 1 (Slot 11)
Index 83 (zero time)
Block: Transducer Block 1 (Slot 11)
Index 13 (zero calibration)
Block: Transducer Block 1 (Slot 11)
Index 143, Bit 0x8000
Block: Transducer Block 1 (Slot 11)
Index 141, Bits 0x0100, 0x0200,
Block: Transducer Block 1 (Slot 11)
Index 12 (zero point)
Zero in progress
Zeroing success
0x0400, and 0x0800
Troubleshoot
Model 2700 Transmitter with PROFIBUS-PA