Calibration
Figure 3-11
Zeroing procedure – ProLink II
Red
Troubleshoot
3.7
Performing density calibration
Density calibration includes the following calibration points:
•
All sensors:
-
D1 calibration (low-density)
-
D2 calibration (high-density)
•
T-Series sensors only:
-
D3 calibration (optional)
-
D4 calibration (optional)
For T-Series sensors, the optional D3 and D4 calibrations could improve the accuracy of the density
measurement. If you choose to perform the D3 and D4 calibrations:
•
Do not perform the D1 or D2 calibrations.
•
Perform the D3 calibration if you have one calibrated fluid.
•
Perform both the D3 and D4 calibrations if you have two calibrated fluids (other than air
and water).
The calibrations that you choose must be performed without interruption, in the order listed here.
Configuration and Use Manual
ProLink >
Calibration >
Zero Calibration
Modify zero time
if required
Perform Auto Zero
Calibration in Progress LED
turns red
Wait until Calibration in
Progress LED turns green
Calibration
Failure LED
Green
Done
39