Emerson Micro Motion 2700 Configuration And Use Manual page 50

With profibus-pa
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Calibration
Figure 3-14
D1 and D2 density calibration – ProLink II
Close shutoff valve
downstream from sensor
42
D1 Calibration
Fill sensor with D1 fluid
ProLink Menu >
Calibration >
Density cal – Point 1
Enter density of D1 fluid
Do Cal
Calibration in Progress light
turns red
Calibration in Progress light
turns green
Close
D2 Calibration
Fill sensor with D2 fluid
ProLink Menu >
Calibration >
Density cal – Point 2
Enter density of D2 fluid
Do Cal
Calibration in Progress light
turns red
Calibration in Progress light
turns green
Close
Done
Model 2700 Transmitter with PROFIBUS-PA

Advertisement

Table of Contents
loading

Table of Contents